PB 840 - Puritan Bennett Ventilator
Reg by: Frances Ellaine M. Patuga
Gas Mixing System
- Range of flow from the mixing system: Can be set to 150 L/min standard temp(STPD).
- Additional flow is available (up to 30 L/min for neonatal circuit type, up to 80 Ltype, and up to 200 L/min for adult circuit type) for compliance compensation.
- Leakage from one gas system to another: Meets standard
- Operating pressure range: 35 to 100 psi (241 to 690 kPa)
- Air/oxygen regulator bleed: Up to 3 L/min
Oxygen and Air Inlet Supplies
- Pressure: 241 to 690 kPa (35 to 100 psi)
- Flow: Maximum of 200 L/minEN ISO13485 :
Max Limited Pressure
-
127.5 cmH2O (125 hPa)
Max Working Pressure
- 100 cmH2O (98.1 hPa), ensured by high pressure limit
- 90 cmH2O (pressure-based ventilation)
Minute Volume (V E TOT ) Capability
-
25 to 75 L/min
Bacteria Filter Efficiency
-
99.97% for nominal particle size of 0.3 μm (micron) at 100 L/min
Monitored Data
-
Breath type, delivered O2, end expiratory pressure, end-inspiratory pressure, exhaled minute volume, exhaled tidal volume, I:E ratio, maximum circuit pressure, mean circuit pressure, spontaneous minute volume,total respiratory rate, rapid shallow breathing index, spontaneous respiratory time.
General
- Modes: Assist/Control, synchronous intermittent mandatory ventilation, or spontaneous
- Respiratory Rates: 1.0 to 100/min
- Inspiratory Pressure: 5 to 90 cm H2O
- Inspiratory Time: 0.2 to 8.0 seconds
- I:E Ratio: ≤ 1:299-4.00:1
- Expiratory Time: Te ≥ 0.2 second
- PEEP: 0 to 45 cm H2O
- Apnea Insp. Pressure: 5 to 90 cm H2O
- Apnea Insp. Time: 0.2 to 8.0 seconds
- Apnea Interval:l 10 to 60 seconds
- Apnea Resp. Rate: 2.0 to 40/min
- Apnea I:E Ratio: ≤ 1.00:1
- Apnea Exp. Time: ≥ 0.2 second